Agilent Cary 500 UV-Vis-NIR Spectrophotometer Moisture Measurement Application Note

This paper presents a method for measuring the moisture content in mineral ore samples using near-infrared diffuse reflectance spectroscopy. It was found that there is a linear relationship between the amount of water absorbed and the non-bonded -OH band at 1900 nm, while no trends were observed for -OH bands at 1400 and 2200 nm. This study is of significant importance for the development of viable methods for extracting, analyzing, and processing oil shale.


Agilent Cary 500 UV-Vis-NIR Spectrophotometer Moisture Measurement Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Product Leaflet
Language
English
Pages
5
File format
PDF
File size
111 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
1341D2EB590915F0D41F614FA3F367C6

About this manual

This document details a study on measuring moisture content in oil shale samples using Near-Infrared (NIR) diffuse reflectance spectroscopy. The research utilized a Cary 500 spectrophotometer equipped with a Praying Mantis diffuse reflectance accessory (DRA) to analyze samples stored at varying humidity levels. The study aimed to establish a correlation between absorbed water and specific spectral bands to predict oil yield from oil shale.

What's inside

  • Introduction
  • Theory
  • Materials and Methods
  • Equipment
  • Protocol
  • Results and Discussion
  • Conclusion
  • References

Specifications

Wavelength range2300 to 1300 nm
Beam modedouble
Average time0.3 sec
Slit heightreduced

Frequently asked questions

Which wavelength band shows a linear relationship with moisture content?

The non-bonded -OH band at 1900 nm shows a linear relationship with the amount of water absorbed.

What equipment is used for the measurement?

The measurement uses a Cary 500 spectrophotometer equipped with a Praying Mantis diffuse reflectance accessory (DRA).

How is the baseline correction performed?

Baseline correction is performed using powdered KBr as a reference.