Agilent UltrAA Lamps Application Note for Arsenic Determination

The product can quickly detect the arsenic content in the ore. This technology uses Agilent UltrAA lamps to allow arsenic to be detected with high levels of other base metals through a flame, without the need to extract arsenic separately or use a different detection technique.


Agilent UltrAA Lamps Application Note for Arsenic Determination - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
59 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
9465EC4FD402706F5FC4F172EF5BE29D

About this manual

Arsenic presents a unique challenge in flame atomic absorption spectroscopy (AAS) due to its low concentration levels, which are difficult to determine alongside high levels of base metals like iron, nickel, and cobalt. Traditional techniques such as hydride generation or graphite atomization are often too slow for rapid screening applications.

The document details a method using Agilent UltrAA boosted discharge hollow cathode lamps combined with a nitrous oxide-acetylene flame. This approach allows for the determination of low-level arsenic in a single extracted solution without separate extraction procedures, utilizing wavelengths at 189.0 nm and 193.7 nm that are not normally available on other systems.

What's inside

  • Introduction
  • Instrumentation
  • Sample Preparation
  • Calibration
  • Results

Frequently asked questions

What is the minimum arsenic concentration detectable in the final solution?

The limits of detection ranged from 0.05 to 0.14 mg/L in final solution.

What is the recommended wavelength for analyzing arsenic with an air-acetylene flame?

The most sensitive conditions are found at 189.0 nm using an air-acetylene flame.

What is the required sample mass for preparing nickel ore samples?

The procedure requires treating a milled ore sample of 1 g with mixed acid reagents.