Agilent J&W FactorFour VF-5ms GC Column Application Note

This application note describes a multi-residue method for polycyclic aromatic hydrocarbons (PAHs), polychlorinated biphenyls (PCBs) and organochlorine pesticides (OCPs). The method was used to analyze 16 PAHs, 17 PCBs and 24 OCPs at different concentration levels in a single run. The results showed that the method is suitable for the analysis of these compounds.


Agilent J&W FactorFour VF-5ms GC Column Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
5
File format
PDF
File size
167 KB
Published
11 June, 2012
Updated
30 September, 2023
Model
VF-5MS
MD5 checksum
BD9041C382D16076B2959214C28400CF

About this manual

The document details an analytical method to analyze 16 polycyclic aromatic hydrocarbons (PAHs), 17 polychlorinated biphenyls (PCBs), and 24 organochlorine pesticides (OCPs) in a single run. This approach is designed to improve productivity for environmental samples by reducing analysis time compared to performing separate analyses for each compound group.

The method utilizes GC/MS with the Agilent J&W FactorFour VF-5ms column, achieving separation of 57 compounds within 32 minutes. The note discusses specific challenges in separating certain peak pairs that have identical masses or difficult resolution using mass spectrometry alone, providing strategies for confirmation via Selected Ion Monitoring (SIM).

What's inside

  • Introduction
  • Conditions
  • Results and Discussion
  • Peak identification

Frequently asked questions

How long does the multi-residue analysis of PAHs, PCBs, and OCPs take on this column?

The VF-5ms column delivers a multi-residue analysis of PAHs, PCBs, and OCPs in 32 minutes.

What are the specified sample concentrations for PAHs, OCPs, and PCBs in this method?

The concentration of the PAHs was ten times higher than the concentration of PCBs and OCPs, with PAHs at 1 wg/mL and OCPs and PCBs at 0.1 yg/mL.

What is the flow rate of the carrier gas used in this analysis?

The carrier gas is Helium set to a constant flow of 1 mL/min.