Agilent SIPS Accessory Application Note: Obtaining Optimum Performance

This document describes the features and performance of the SIPS accessory, as well as the precautions to be taken when using it.


Agilent SIPS Accessory Application Note: Obtaining Optimum Performance - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
30 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
4161C2FACB979100CA5B34F57E7E06BA

About this manual

The document explains that the Agilent SIPS (Single Injection Pumping System) accessory, introduced in 1994, is a dilution system for flame atomic absorption. It addresses the issue of 'spalling,' where composite pump tubing breaks into particles that can block the nebulizer.

To achieve reliable operation and optimum performance, the text details four key procedures: using only Agilent-supplied SIPS pump tubing, determining the correct arm pressure for each unit, conditioning new or used pump tubes with detergent solutions, and adding Triton X-100 to diluent solutions.

What's inside

  • Introduction
  • The Effect of Spalling
  • Why Use Composite Materials?
  • Achieving Reliable SIPS Operation
  • Use Only Agilent-Supplied SIPS Pump Tubing
  • Determine the Correct Arm Pressure
  • Condition the Pump Tubing
  • Add a Detergent

Specifications

Introduction DateDecember 1994
Recommended Detergent Concentration for Conditioning (New)1% solution (mass/volume) of Triton X-100
Recommended Detergent Concentration for Conditioning (Used)0.01 % Triton X-100

Frequently asked questions

What is spalling in the context of SIPS pump tubing?

Spalling is a variable effect where very small particles of the composite tube material break away and can stick together to cause blockage of the nebulizer.

Why are very dilute solutions more likely to induce spalling?

The document states that it has been found that very dilute solutions are more likely to induce spalling and block the nebulizer than concentrated solutions.

When should the arm pressure setting for a SIPS unit be recalibrated?

The procedure need only be repeated if the SIPS unit is repaired or changed, such as upgrading from a SIPS-10 to a dual pump SIPS-20.