About this manual

The document details a colorimetric method for determining dissolved hexavalent chromium content in water. The procedure involves reacting the sample with 1,5-diphenylcarbazide in an acid medium to develop a reddish-purple color, which is then measured at 540 nm using the DMS 90 spectrophotometer.

What's inside

  • Apparatus
  • Reagents
  • Experimental Procedure
  • Color development
  • Results
  • Interferences

Frequently asked questions

What wavelength is used to measure hexavalent chromium?

The absorbance of the developed color is measured at 540 nm.

What slit width was used for this experiment?

A slit width of 2 nm was used because the chromium peak is broad.

How long should the solution stand for color development?

The test solution should be allowed to stand for 10 minutes to ensure consistent results.

Which elements interfere with the chromium determination?

Iron (III), vanadium, molybdenum, mercury, and nitrites have adverse effects on the color development.