Varian DMS 90 Spectrophotometer Protein Determination Guide

This manual introduces how to use Varian DMS 90 spectrophotometer to determine protein concentration.


Varian DMS 90 Spectrophotometer Protein Determination Guide - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
5
File format
PDF
File size
565 KB
Published
11 June, 2012
Updated
01 October, 2023
Model
DMS 90
MD5 checksum
A1C4FA997ED33830ACF64EE8A9914E9A

About this manual

This document describes various laboratory procedures for determining protein concentration in solutions. It discusses the principles of measuring absorbance at 280 nm and compares this with other methods such as the biuret reaction, the Lowry assay, and the Coomassie Brilliant Blue G-250 dye-binding procedure.

What's inside

  • Introduction
  • Reagents and Materials
  • Equipment
  • Procedure: 280 nm Absorbance
  • Procedure: Biuret Method
  • Procedure: Lowry Assay
  • Procedure: Coomassie Blue Assay
  • Results and Discussion

Specifications

Spectral Bandwidth (280 nm)1.0 nm
Wavelength for Biuret Assay540 nm
Sensitivity (280 nm for Albumin)0.64 absorbance units per mg/mL
Sensitivity (Lowry Assay)1.67 absorbance units per mg/mL
Wavelength for Coomassie Assay595 nm

Frequently asked questions

What is the sensitivity of the 280 nm assay for bovine plasma albumin?

The sensitivity is 0.64 absorbance units per mg/mL.

How much more sensitive is the Lowry assay compared to 280 nm measurement?

It is approximately three times more sensitive.

What wavelength should be used for the Coomassie Brilliant Blue G-250 dye-binding procedure?

The absorbance should be measured at 595 nm.