Agilent 971-FP Flash Purification of Non-UV-Active Compounds Application Note

This document demonstrates the benefits of using an evaporative light scattering detector (ELSD) coupled to the Agilent 971-FP Flash instrument for the analysis of a sample containing a mixture of UV and non-UV-active compounds. The ELSD can detect compounds without UV chromophores, preventing the loss of these compounds that would go undetected by UV and be discarded.


Agilent 971-FP Flash Purification of Non-UV-Active Compounds Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
333 KB
Published
11 June, 2012
Updated
28 September, 2023
Model
971-FP
MD5 checksum
13A918D76455187DD32532370F2088F2

About this manual

This document describes a method for flash purification of compounds that lack UV chromophores using an Evaporative Light Scattering Detector (ELSD). It highlights the benefits of coupling an Agilent 385-ELSD with an Agilent 971-FP Flash instrument to prevent the loss of sample components that are undetected by UV alone.

The note details a specific experiment using a mixture of B-carotene, tocopherols, and aminophenols. It explains how the software allows direct control of the ELSD settings and how fraction collection can be triggered by both UV and ELSD responses to ensure all components are collected.

What's inside

  • Introduction
  • Software Control
  • Methods and Materials
  • Results and Discussion
  • Conclusion

Specifications

ColumnSF15-12g Si 35
Eluent AHexane
Eluent BEthyl acetate
Flow Rate18 mL/min
Injection Volume0.5mL
TemperatureAmbient
UV Detection Wavelength254 nm
ELSD Nebulizer Temperature15°C
ELSD Evaporation Temperature30 °C
ELSD Gas Flow1.2 SLM

Frequently asked questions

What is the advantage of using an ELSD with a Flash instrument for non-UV-active compounds?

It prevents the loss of sample components that remain undetected by UV and are therefore not isolated by the fraction collector.

How does the Agilent 971-FP software control the ELSD?

The software permits direct control of the ELSD via an on-screen user interface, allowing the operator to change detection settings as required.

How is fraction collection triggered in this method?

Fraction collection is triggered by both the UV and ELSD responses to ensure all components are collected.