Agilent 490 Micro GC C1-C3 Hydrocarbon Analysis Application Note

This application note shows the possibilities and limitations in fast analysis of saturated and unsaturated C1 to C3 hydrocarbons using an Agilent 490 Micro GC. The chromatograms and results outline the similarities and differences when using a PoraPLOT U and a PoraPLOT Q column channels.


Agilent 490 Micro GC C1-C3 Hydrocarbon Analysis Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
320 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
FF2185E42692F9995B2809AD9F22028A

About this manual

This document describes the possibilities and limitations in fast analysis of saturated and unsaturated C1 to C3 hydrocarbons using an Agilent 490 Micro GC. It outlines the separation characteristics when using a PoraPLOT U column channel versus a PoraPLOT Q column channel, noting that both are capable of resolving methane from the composite air peak and separating CO from methane and C2 hydrocarbons.

What's inside

  • Introduction
  • Separation Characteristics
  • Instrumentation
  • Setup Parameters
  • Sample Information

Frequently asked questions

What hydrocarbons can the PoraPLOT U column channel separate from each other?

The PoraPLOT U column channel provides baseline separation for ethane, ethylene, and acetylene.

Which hydrocarbons coelute when using the PoraPLOT O column channel?

When using the PoraPLOT O column channel, ethylene and acetylene coelute.

What is the column temperature used in the described setup?

The column temperature used in the described setup is 80°C.