Agilent Bio SEC-3 Application Note for rEPO Separation

This document describes the separation of recombinant human erythropoietin (rEPO) using Agilent Bio SEC-3. The method utilizes Agilent Bio SEC columns, which are designed to improve separation efficiency and stability by reducing secondary interactions. The columns are optimized for use with organic modifier denaturants and low salt concentrations, making them ideal for quality control in protein primary structure analysis.


Agilent Bio SEC-3 Application Note for rEPO Separation - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
891 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
84D75AF34F64FAC44E1EA19935B28FD5

About this manual

The document details a method for monitoring the purity and stability of recombinant human Erythropoietin (rEPO). It utilizes size exclusion chromatography with Agilent Bio SEC-3 columns to separate intact rEPO from aggregates, dimers, and monomers. The study also evaluates the separation performance under stress conditions, specifically heat-stressed samples at low pH and neutral pH.

What's inside

  • Introduction
  • Material and Methods
  • HPLC system
  • HPLC column
  • LC methods
  • Results and Discussion
  • Separation of intact rEPO
  • Separation of heat-stressed rEPO
  • Conclusion

Specifications

Column Part Number5190-250
Column Dimensions4.6 x 300 mm
Particle Size3 um
Flow Rate0.35 mL/min
Injection Volume5 ul
Mobile Phase150 mM sodium phosphate buffer, pH 7.0
Detector Wavelength225 nm

Frequently asked questions

What system was used for the rEPO separation?

The Agilent 1260 Infinity Bio-inert Quaternary LC system.

What is the recommended flow rate for intact rEPO separation?

The optimized flow rate was 0.35 mL/min.

Download PDF (891 KB)