Agilent Bond Elut QuEChERS AOAC Kit Application Note for Pesticide Residue Analysis

Analysis of Pesticide Residues in Apples using Agilent Bond Elut QuEChERS AOAC Kit by LC/MS/MS Detection


Agilent Bond Elut QuEChERS AOAC Kit Application Note for Pesticide Residue Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
14
File format
PDF
File size
683 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
A25894EF157BAFCEC198873DF9A5B089

About this manual

This document details a validated method for extracting and cleaning up 16 pesticide residues from apple samples. The procedure utilizes the Agilent Bond Elut QuEChERS AOAC Extraction kit (p/n 5982-5755) followed by cleanup with Bond Elut QuEChERS AOAC dispersive SPE kits (p/n 5982-5022 and 5982-5058). The prepared extracts are analyzed using an Agilent 1200 HPLC coupled to a 6410 Triple Quadrupole LC/MS/MS system.

What's inside

  • Abstract
  • Introduction
  • Reagents and Chemicals
  • Solutions and Standards
  • Equipment and Material
  • Instrument Condition
  • Sample preparation

Frequently asked questions

What is the limit of quantitation for pesticides in apple using this method?

The limit of quantitation is 5 ng/g.

What spiking levels were used for the recovery experiments?

The spiking levels were 10, 50, and 200 ng/g.

What is the average mean recovery percentage for the pesticides tested?

The average mean recovery is 95.4%.