Agilent 7693A ALS Automated Sample Preparation for Fatty Acids Application Note

A miniaturized automated sample preparation method is described for the determination of fatty acid methyl esters in blood or plasma samples using the Agilent 7693 Series Automatic Liquid Sampler. Fifty-microliter samples are derivatized using acetyl chloride in methanol and the fatty acid methyl esters are extracted in iso-octane. Extracts are analyzed by GC-FID. Excellent repeatability is obtained and the method can be applied for fully automated fatty acid profiling in blood and plasma samples during dietary, biomedical or metabolomic studies.


Agilent 7693A ALS Automated Sample Preparation for Fatty Acids Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
266 KB
Published
11 June, 2012
Updated
29 September, 2023
Model
7693A
MD5 checksum
6FE29A15C74A2D2378452F527D75257E

About this manual

The document details a fully automated sample preparation method for determining fatty acid methyl esters (FAMEs) in blood or plasma samples using the Agilent 7693 Series Automatic Liquid Sampler (ALS). The method involves derivatizing 50-microliter samples with acetyl chloride in methanol and extracting the fatty acid methyl esters in iso-octane for analysis by GC-FID.

The procedure utilizes a specific sequence programmed in GC ChemStation to handle reagents, heating, mixing, and extraction steps. The text provides a detailed sample preparation sequence, analytical conditions for the Agilent 7890 GC System, and results demonstrating excellent reproducibility with an average RSD of 2% for the analyzed fatty acids.

What's inside

  • Introduction
  • Experimental
  • Sample Preparation Sequence
  • Analytical Conditions
  • Results and discussion
  • Peak Identification
  • Reproducibility

Specifications

Sample Volume50 microliters
Derivatization Reagent5% acetyl chloride in methanol
Extraction Solventiso-octane
Heating Temperature75 °C
Heating Duration600 seconds
Mixing Speed4000 rpm
Column TypeHP-88
Inlet Temperature250 °C
Total Run Time84.33 min

Frequently asked questions

What is the sample volume required for this method?

The method uses fifty-microliter samples of blood or plasma.

What is the heating temperature and duration for the derivatization step?

The sample is heated and mixed at 75°C for 600 seconds (10 minutes).

Which column is recommended for optimal resolution of cis-trans fatty acids?

A long polar Agilent HP-88 GC column is used for optimal resolution.