Start here

Winsen MC119 Catalytic Combustion Gas Sensor User Manual

Winsen MC119 Catalytic Combustion Gas Sensor User Manual

User Manual 7 pages · 504 KB · ZH

MC119 is a catalytic combustion gas sensor from Zhengzhou Winsen Electronic Technology Co., Ltd. It works by catalytic combustion: a detection element and a compensation element form one arm of a bridge, and the bridge output voltage changes as combustible gas concentration rises. The compensation element provides reference and temperature/humidity compensation.

The manual gives the sensor's technical specifications and performance curves, and lists usage precautions. Its main applications include concentration detection of hydrogen, acetylene, gasoline, and organic vapors such as alcohols, ketones, and benzene at industrial sites. This manual is written in Chinese.

Read online

Winsen MC119 specifications

Product ModelMC119
Product TypeCatalytic combustion gas sensor
Standard PackageMetal package, metallurgical powder mesh
Operating Voltage2.5 ± 0.1 V
Operating Current100 ± 10 mA
Sensitivity (1% H2)20–40 mV
Sensitivity (1% Acetylene)40–90 mV
Measuring Range0–100 %LEL
Response Time (90%)≤10 s
Recovery Time (90%)≤30 s
Operating Environment-40 to +70 °C, below 95% RH
Storage Environment-20 to +70 °C, below 95% RH
Dimensions (L × W × H)19 × 9.5 × 14 mm
Explosion-Proof MarkingExdibⅠ

Source: Winsen MC119 Catalytic Combustion Gas Sensor User Manual

All Winsen MC119 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the measuring range of the MC119?

The manual states a measuring range of 0 to 100 %LEL.

From: Winsen MC119 Catalytic Combustion Gas Sensor User Manual

What is the response time of the MC119?

The response time (90%) is 10 seconds or less.

From: Winsen MC119 Catalytic Combustion Gas Sensor User Manual

What gases can the MC119 sensor detect?

It is used in industrial settings to detect the concentration of hydrogen, ethyne, gasoline and VOCs such as alcohol, ketone and benzene, and has good sensitivity to methane.

From: Winsen MC119 Catalytic Flammable Gas Sensor Manual

How should the MC119 sensor be soldered?

Handmade welding is optimal, using rosin soldering flux with least chlorine, a 250 °C soldering iron for less than 3 seconds. For wave-soldering, the speed should be 1–2 m/min with a warm-up temperature of 100±20 °C and welding temperature of 250±10 °C.

From: Winsen MC119 Catalytic Flammable Gas Sensor Manual