OHAUS MB27 Moisture Analyzer User Manual
The document serves as a comprehensive user guide for the OHAUS MB27 Moisture Analyzer. It covers essential safety precautions, including warnings about heat, flammable materials, and proper ventilation requirements.
Installation instructions detail the package contents, location selection criteria, component installation (Draft Shield, Pan Support), and power connection procedures. The manual specifies that the unit requires a warm-up period of at least 15 minutes after connecting to AC power.
OHAUS MB27 specifications
| Drying temperature range | 50° to 160°C in 5° steps |
|---|---|
| Drying time range | 1 to 60 minutes in 30 second steps, 61 to 99 minutes in 1 minute steps |
| Minimum sample weight | greater than 0.5 g |
| Result units | grams, % Moisture, % Solids |
| Warm-up time | at least 15 minutes |
| Data interface | RS232 |
| Power connection | 3-pin power cable with ground connection |
All OHAUS MB27 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How long should the MB27 warm up before use?
After connecting the analyzer to the AC supply in standby mode, allow it to warm up for at least 15 minutes for best results.
What is the minimum sample size for a moisture test?
The sample must be greater than 0.5 g and should be spread evenly across the test pan.
What is the moisture analyzer intended to be used for?
It is to be used exclusively for determining moisture in samples.
What is the minimum sample size for a test?
The sample must be greater than 0.5 g.
What is the minimum sample weight required to initiate a test?
The analyzer initiates a test only when the sample weight is greater than 0.5 g.
How long must the analyzer warm up after connecting to power?
The unit must warm up for at least 15 minutes to achieve best results.