The MB27 Moisture Analyzer is a versatile instrument that can quickly and accurately measure moisture content. It features:
This instruction manual describes the OHAUS MB27 moisture analyzer, a benchtop instrument used for the determination of moisture in samples by heating. It covers safety precautions, package contents, installation and leveling, and explains the display, controls and operating modes including Standby, Weigh, Edit, Run and Result modes.
The manual explains how to prepare and perform a moisture determination test: setting the drying temperature (50° to 160°C in 5° steps) and drying time (timed duration or AUTO), taring the sample pan, and starting the test with a sample greater than 0.5 g. Results can be shown in grams, percent moisture or percent solids, and sent to a printer or computer over RS232. Sections on test optimization, calibration and maintenance are also referenced.
| Drying temperature range | 50° to 160°C in 5° steps |
|---|---|
| Drying time range | 1 to 60 minutes in 30 second steps, 61 to 99 minutes in 1 minute steps |
| Minimum sample weight | greater than 0.5 g |
| Result units | grams, % Moisture, % Solids |
| Warm-up time | at least 15 minutes |
| Data interface | RS232 |
| Power connection | 3-pin power cable with ground connection |
After connecting the analyzer to the AC supply in standby mode, allow it to warm up for at least 15 minutes for best results.
The sample must be greater than 0.5 g and should be spread evenly across the test pan.
The drying temperature can be set from 50° to 160°C in 5° steps.
Yes, the time parameter can be set to AUTO, where the test ends when the analyzer detects that the weight loss has ended.