Key facts
- Brand
- Fluke
- Model
- 5700A/5720A
- Category
- Other Electronics
- Industry
- Industrial
- Documents
- 4
- Languages
- EN · ZH
- Last added
- 2026-11-22
Start here

Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual
This manual describes the Fluke 5700A/5720A Series II calibrator, a precision multi‑function source for DC/AC voltage, current, and resistance. It covers installation, front‑panel operation, remote control, and detailed calibration procedures.
Safety warnings, warranty information, specifications, and interface options such as IEEE‑488 (GPIB) and RS‑232C are included, along with troubleshooting and service guidance.
Fluke 5700A/5720A specifications
| DC Voltage Output | up to 1100 V |
|---|---|
| AC Voltage Output | up to 1100 V, 10 Hz‑1.2 MHz |
| Current Output | AC/DC up to 2.2 A, 10 Hz‑10 kHz |
| Resistance Range | 1 Ω to 100 MΩ |
| Optional Wideband AC Voltage | 300 µV‑3.5 V into 50 Ω, –57 dBm to +24 dBm, 10 Hz‑30 MHz |
Source: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual
All Fluke 5700A/5720A documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is Artifact Calibration?
It is a process that uses only three artifact standards (10V dc, 1Ω, and 10 kΩ resistance) to calibrate all ranges and functions, reducing the process from several hours to about one hour.
From: Fluke 5700A/5720A Multifunction Calibrator Product Leaflet
How does the Cal Check function work?
The automated Cal Check function verifies each range and function against internal standards to ensure they remain within specification between calibrations.
From: Fluke 5700A/5720A Multifunction Calibrator Product Leaflet
How long should the calibrator warm up after being off for a long time?
Warm it up for 30 minutes.
From: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual
How often must DC zero calibration be performed according to specification?
At least every 30 days.
From: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual
What is the maximum voltage that can appear on the output terminals?
Lethal voltages of up to 1100 V AC or DC may be present on the output terminals.
From: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual
How long is the warranty period for the calibrator?
The warranty period is one year from the date of shipment.
From: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual