Fluke 5700A/5720A Manuals

ManualLib has 4 documents for the Fluke 5700A/5720A (Other Electronics): 1 product leaflet, 3 user manuals. Every file can be read online or downloaded as a PDF for free.

Product Leaflet 1User Manual 3

Key facts

Brand
Fluke
Model
5700A/5720A
Industry
Industrial
Documents
4
Languages
EN · ZH
Last added
2026-11-22

Start here

Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual

Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual

User Manual 332 pages · 3.3 MB · EN

This manual describes the Fluke 5700A/5720A Series II calibrator, a precision multi‑function source for DC/AC voltage, current, and resistance. It covers installation, front‑panel operation, remote control, and detailed calibration procedures.

Safety warnings, warranty information, specifications, and interface options such as IEEE‑488 (GPIB) and RS‑232C are included, along with troubleshooting and service guidance.

Read online

Fluke 5700A/5720A specifications

DC Voltage Outputup to 1100 V
AC Voltage Outputup to 1100 V, 10 Hz‑1.2 MHz
Current OutputAC/DC up to 2.2 A, 10 Hz‑10 kHz
Resistance Range1 Ω to 100 MΩ
Optional Wideband AC Voltage300 µV‑3.5 V into 50 Ω, –57 dBm to +24 dBm, 10 Hz‑30 MHz

Source: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual

All Fluke 5700A/5720A documents by type

Questions answered in these manuals

What is Artifact Calibration?

It is a process that uses only three artifact standards (10V dc, 1Ω, and 10 kΩ resistance) to calibrate all ranges and functions, reducing the process from several hours to about one hour.

From: Fluke 5700A/5720A Multifunction Calibrator Product Leaflet

How does the Cal Check function work?

The automated Cal Check function verifies each range and function against internal standards to ensure they remain within specification between calibrations.

From: Fluke 5700A/5720A Multifunction Calibrator Product Leaflet

How long should the calibrator warm up after being off for a long time?

Warm it up for 30 minutes.

From: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual

How often must DC zero calibration be performed according to specification?

At least every 30 days.

From: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual

What is the maximum voltage that can appear on the output terminals?

Lethal voltages of up to 1100 V AC or DC may be present on the output terminals.

From: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual

How long is the warranty period for the calibrator?

The warranty period is one year from the date of shipment.

From: Fluke 5700A/5720A Series II Multi-Function Calibrator User Manual