Key facts

Brand
Fluke
Model
TiS10
Documents
3
Languages
EN · ES
Last added
2023-11-15

Start here

Fluke TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 Performance Series Thermal Imagers Users Manual

Fluke TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 Performance Series Thermal Imagers Users Manual

User Manual 52 pages · 662 KB · EN

This user manual covers Fluke TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, and TiS65 Performance Series thermal imagers. It describes safety information, accessories, features and controls, image capture, temperature measurement, menus, maintenance, and specifications.

The manual includes general and detailed specifications for the imager models, including temperature ranges, imaging performance, battery and power information, and supported languages.

Read online

Fluke TiS10 specifications

Operating temperature-10 °C to +50 °C (14 °F to 122 °F)
Storage temperature-20 °C to +50 °C (-4 °F to 122 °F)
Relative humidity10 % to 95 % non-condensing
Operating altitude2,000 m (6,562 ft)
Storage altitude12,000 m (39,370 ft)
Display8.9 cm (3.5 in) diagonal landscape color VGA (320 x 240) LCD with backlight
SoftwareSmartView full analysis and reporting software available for free download at www.fluke.com
BatteriesSBP3 Lithium-ion rechargeable smart battery pack
Battery life4 hours continuous use (assumes 50 % brightness of LCD)
Battery charge time2.5 hours to full charge
AC battery chargeTi SBC3B Two Bay Battery Charger (100 V ac to 240 V ac, 50/60 Hz) or in-Imager charging; optional 12 V automotive charging adapter
AC operation100 V ac to 240 V ac, 50/60 Hz, ac mains adapters included
Power saveUser-selectable Sleep and Power Off modes
Safety standardsMains IEC 61010-1, Overvoltage Category II, Pollution Degree 2; IEC 60825-1, Class 2, <1 mW
EMCIEC 61326-1: Basic Electromagnetic Environment; CISPR 11: Group 1, Class A
Korea (KCC)Class A Equipment (Industrial Broadcasting & Communication Equipment)
USA (FCC)47 CFR 15 subpart B; exempt device per clause 15.103
Vibration2 G, IEC 68-2-6
Shock25 G, IEC 68-2-29
Drop2 m (6.5 ft) with standard lens
Size (H x W x L)26.7 cm x 10.1 cm x 14.5 cm (10.5 in x 4.0 in x 5.7 in)
Weight (TiS10, TiS20, TiS40, TiS50, TiS60)0.72 kg (1.6 lb)
Weight (TiS45, TiS55, TiS65)0.77 kg (1.7 lb)
Enclosure ratingIP54
Warranty2 years
Calibration cycle2 years (assumes normal operation and normal aging)
Supported languagesCzech, Dutch, English, Finnish, French, German, Hungarian, Italian, Japanese, Korean, Polish, Portuguese, Russian, Simplified Chinese, Spanish, Swedish, Traditional Chinese, and Turkish
Temperature range (TiS10)-20 °C to +250 °C (-4 °F to +482 °F)
Temperature range (TiS20, TiS40, TiS45)-20 °C to +350 °C (-4 °F to +662 °F)
Temperature range (TiS50, TiS55)-20 °C to +450 °C (-4 °F to +842 °F)
Temperature range (TiS60, TiS65)-20 °C to +550 °C (-4 °F to +1022 °F)
Temperature accuracy±2 °C or 2 % (whichever is greater) at 25 °C ambient
On-screen emissivity correction1 % to 100 %
On-screen reflected background temperature compensationYes, default is 22 °C
On-screen transmission correction (TiS45, TiS55, TiS65)1 % to 100 %
IFOV (TiS10)7.8 mRad, D:S 137:1
IFOV (TiS20)5.2 mRad, D:S 206:1
IFOV (TiS40, TiS45)3.9 mRad, D:S 275:1
IFOV (TiS50, TiS55)2.8 mRad, D:S 377:1
IFOV (TiS60, TiS65)2.4 mRad, D:S 446:1
Field of view35.7 ° x 26.8 °
Image capture frequency9 Hz or 30 Hz refresh rate
Detector typeFocal Plane Array, uncooled microbolometer
Thermal sensitivity (TiS10)≤150 mK
Thermal sensitivity (TiS20)≤100 mK
Thermal sensitivity (TiS40, TiS45)≤90 mK
Thermal sensitivity (TiS50, TiS55, TiS60, TiS65)≤80 mK
Total pixels (TiS10)80 x 60
Total pixels (TiS20)120 x 90
Total pixels (TiS40, TiS45)160 x 120
Total pixels (TiS50, TiS55)220 x 165
Total pixels (TiS60, TiS65)260 x 195
Infrared spectral band7.5 μm to 14 μm (long wave)
Visual cameraIndustrial performance 5.0 megapixel
Minimum parallax alignment~60 cm (~24 in) with standard IR lens
Focus mechanismFixed Focus: TiS10, TiS20, TiS40, TiS50, TiS60; Manual Focus: TiS45, TiS55, TiS65

Source: Fluke TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 Performance Series Thermal Imagers Users Manual

All Fluke TiS10 documents by type

Questions answered in these manuals

What is the warranty period for these thermal imagers?

The warranty period is two years and begins on the date of shipment.

From: Fluke TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 Performance Series Thermal Imagers Users Manual

How long does the battery last during continuous use?

Battery life is 4 hours continuous use, assuming 50 % brightness of the LCD.

From: Fluke TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 Performance Series Thermal Imagers Users Manual

What temperature range do the TiS60 and TiS65 support?

They measure from -20 °C to +550 °C (-4 °F to +1022 °F).

From: Fluke TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 Performance Series Thermal Imagers User Manual

How accurate is the temperature measurement?

±2 °C or 2%, whichever is greater, at 25 °C ambient temperature.

From: Fluke TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 Performance Series Thermal Imagers User Manual

What is the effective spectral response for the TiS45, TiS55, and TiS65 models?

The effective spectral response for these models is shown in Figure A and requires temperature correction values from Table A when emissivity is not set to 1.0.

From: TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 User, Web Only & TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 Safety Information Manual Supplement

What is the default SSID for the imager's WiFi hotspot?

The default SSID is 'Fluke-Camera' and the password is disabled for new imagers.

From: TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 User, Web Only & TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 Safety Information Manual Supplement