Fluke 9500B Manuals

ManualLib has 2 documents for the Fluke 9500B: 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Brand
Fluke
Model
9500B
Documents
2
Languages
EN · ZH
Last added
2023-10-18

Start here

User ManualFluke 9500B Oscilloscope Calibrator User Manual

Fluke 9500B Oscilloscope Calibrator User Manual

User Manual 12 pages · 627 KB · ZH

The Fluke 9500B is a high-performance, fully automated oscilloscope calibrator designed to calibrate modern multi-channel oscilloscopes without manual intervention. It features active signal heads that generate calibration signals directly at the oscilloscope input, eliminating signal distortion caused by cables or switches. The system supports upgrades from 600MHz to 6.4GHz, with optional active signal heads for enhanced performance.

It integrates with MET/CAL software for automated calibration, report generation, and inventory management, supporting multiple oscilloscope models and offering future-proofing through software updates and support plans. This manual is written in Chinese.

Read online

Fluke 9500B specifications

Frequency Uncertainty>15kHz ±0.25ppm ±12mHz, ≤15kHz ±3ppm
Oscilloscope Bandwidth Requirement≥6GHz for Risetime measurements (≥20GHz for 9550 and 9560)
Trigger Cable RequirementShort (no longer than 19.5 inches or 0.5 m), high quality, trigger cable fitted with SMA connectors

Source: Fluke 9500B Oscilloscope Calibrator User Manual Supplement

All Fluke 9500B documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What equipment is required to calibrate the 9500B Oscilloscope Calibrator?

A verified Model 9500B Mainframe, a high-bandwidth sampling oscilloscope with bandwidth ≥6GHz (≥20GHz for 9550 and 9560), a 50e SMA–SMA co-axial ‘Trigger’ cable, and possibly a high-bandwidth coaxial attenuator.

From: Fluke 9500B Oscilloscope Calibrator User Manual Supplement

How do you calibrate the 25ps edge risetime of the 9550?

Connect the 9550 to a high-bandwidth oscilloscope, set the oscilloscope to average 256 samples, and measure the rising and falling edge transition times using the 9500B’s 25ps FAST Edge function. Calculate the average of 8 measurements and use the provided formula to determine the calibrated edge rise time.

From: Fluke 9500B Oscilloscope Calibrator User Manual Supplement

How does the 9500B eliminate signal distortion during calibration?

It uses active signal head technology to generate calibration signals directly at the oscilloscope input, eliminating distortion caused by cables or switches.

From: Fluke 9500B Oscilloscope Calibrator User Manual

What is the maximum frequency of the sine wave output?

Up to 6.4GHz, depending on the active signal head model.

From: Fluke 9500B Oscilloscope Calibrator User Manual