Fluke 6331 Manuals

ManualLib has 2 documents for the Fluke 6331 (Calibration Components): 1 other, 1 product leaflet. Every file can be read online or downloaded as a PDF for free.

Other 1Product Leaflet 1

Key facts

Brand
Fluke
Model
6331
Industry
Industrial
Documents
2
Languages
EN · ZH
Last added
2026-11-18

Start here

Fluke 6331 User Guide Supplement

Fluke 6331 User Guide Supplement

Other 3 pages · 182 KB · EN

This is a manual supplement for the Fluke 6331, issued to keep the 6331 Users Guide accurate. It is dated 3/21 and carries Issue 4 of the supplement; the original guide was printed in January 2013 and distributed as an electronic manual on CD-ROM (CD PN: 3750900). The supplement lists numbered changes that add to, replace or delete passages on specific pages of the guide.

The changes cover a note on control sensitivity to electromagnetic and conducted interference in the Environmental Conditions section, a new procedure for installing the unit on a concrete floor with a bracket and anchor bolts, a deletion in the Unpacking section, a rewritten Reset Cutout and Cutout description, removal of the cutout reset mode parameter and its menu entries, a replacement for section 9.16.1 CTO, a replacement maintenance bullet on checking the over-temperature cutout, and a replacement troubleshooting bullet for a flashing "Cut-out" display.

Read online

Fluke 6331 specifications

Radiated EM disturbance frequency range250 MHz to 400 MHz
Stability change for radiated disturbance above 1 V/m to 3 V/mAdd 0.0025 °C to the stability specification
Conducted disturbance frequency range8 MHz to 80 MHz
Stability change for conducted disturbance above 3 VAdd 0.005 °C to the stability specification
Floor hole size for bracket installation1/4 inch x 1 3/4 inch deep (approximately 6.5 mm x 32 mm)
Air circulation clearanceMinimum of 6 inches
Cutout set-point adjustment10 °C or 20 °C above the maximum bath operating temperature
Over-temperature cutout check intervalEvery 6 months

Source: Fluke 6331 User Guide Supplement

All Fluke 6331 documents by type

Questions answered in these manuals

How often should the over-temperature cutout be checked?

It should be checked every 6 months to see that it is working properly.

From: Fluke 6331 User Guide Supplement

What happens if the cutout is activated by excessive bath temperature?

Power to the heater is shut off, the bath cools, and the display flashes Cut-out until the temperature falls below the reset temperature and the cutout is reset.

From: Fluke 6331 User Guide Supplement

What is the temperature range of the 6331 bath?

The 6331 deep-well bath covers a temperature range of 35°C to 300°C.

From: Fluke CPTB6331 Temperature Verification System Product Leaflet

Can the system operate independently without auxiliary equipment?

Yes, every device in the Fluke system can operate and control independently without needing other auxiliary equipment.

From: Fluke CPTB6331 Temperature Verification System Product Leaflet