Start here

Fluke 6331 User Guide Supplement
This is a manual supplement for the Fluke 6331, issued to keep the 6331 Users Guide accurate. It is dated 3/21 and carries Issue 4 of the supplement; the original guide was printed in January 2013 and distributed as an electronic manual on CD-ROM (CD PN: 3750900). The supplement lists numbered changes that add to, replace or delete passages on specific pages of the guide.
The changes cover a note on control sensitivity to electromagnetic and conducted interference in the Environmental Conditions section, a new procedure for installing the unit on a concrete floor with a bracket and anchor bolts, a deletion in the Unpacking section, a rewritten Reset Cutout and Cutout description, removal of the cutout reset mode parameter and its menu entries, a replacement for section 9.16.1 CTO, a replacement maintenance bullet on checking the over-temperature cutout, and a replacement troubleshooting bullet for a flashing "Cut-out" display.
Fluke 6331 specifications
| Radiated EM disturbance frequency range | 250 MHz to 400 MHz |
|---|---|
| Stability change for radiated disturbance above 1 V/m to 3 V/m | Add 0.0025 °C to the stability specification |
| Conducted disturbance frequency range | 8 MHz to 80 MHz |
| Stability change for conducted disturbance above 3 V | Add 0.005 °C to the stability specification |
| Floor hole size for bracket installation | 1/4 inch x 1 3/4 inch deep (approximately 6.5 mm x 32 mm) |
| Air circulation clearance | Minimum of 6 inches |
| Cutout set-point adjustment | 10 °C or 20 °C above the maximum bath operating temperature |
| Over-temperature cutout check interval | Every 6 months |
Source: Fluke 6331 User Guide Supplement
All Fluke 6331 documents by type
Questions answered in these manuals
How often should the over-temperature cutout be checked?
It should be checked every 6 months to see that it is working properly.
What happens if the cutout is activated by excessive bath temperature?
Power to the heater is shut off, the bath cools, and the display flashes Cut-out until the temperature falls below the reset temperature and the cutout is reset.
What is the temperature range of the 6331 bath?
The 6331 deep-well bath covers a temperature range of 35°C to 300°C.
From: Fluke CPTB6331 Temperature Verification System Product Leaflet
Can the system operate independently without auxiliary equipment?
Yes, every device in the Fluke system can operate and control independently without needing other auxiliary equipment.
From: Fluke CPTB6331 Temperature Verification System Product Leaflet