Key facts
- Brand
- Fluke
- Model
- 5626/5628
- Category
- Temperature sensors
- Industry
- Industrial
- Documents
- 2
- Languages
- EN
- Last added
- 2026-11-06
Start here

Fluke 5626/5628 Platinum Resistance Thermometer User's Guide
This user's guide covers the Fluke 5626 and 5628 Platinum Resistance Thermometers (PRT), secondary standard instruments that convert temperature to resistance. The 5626 has a resistance of 100 Ω and the 5628 has 25.5 Ω; both cover the ITS-90 range from −200 °C to 661 °C and are used with a readout device to detect temperature changes or actual temperature. The document was published in March 2014.
The guide opens with symbols and safety information, then gives specifications, construction and electrical circuit details. It continues with installation (environmental issues, mounting, lead wire identification), PRT care and handling guidelines, operation (comparison calibration, immersion requirements, thermal EMF), accessories such as case options and PRT termination, and a troubleshooting chapter. The PRTs are calibrated in Hart's Scientific Calibration laboratory and supplied with a NIST traceable Report of Calibration.
Fluke 5626/5628 specifications
| Temperature range | −200 °C to 661 °C |
|---|---|
| Nominal resistance, 5626 | 100 Ω ±1 Ω at 0.01 °C |
| Nominal resistance, 5628 | 25.5 Ω ±0.5 Ω at 0.01 °C |
| Calibration uncertainty (k=2) | ±0.006 °C at −200 °C; ±0.004 °C at 0 °C; ±0.009 °C at 420 °C; ±0.014 °C at 661 °C |
| Short-term stability | ±0.003 °C (5626), ±0.002 °C (5628) |
| Long-term drift (Rtp) | ≤ 0.03 °C (5626) / ≤ 0.02 °C (5628) with 500 hours at 661 °C |
| Sheath | Inconel 600, 6.35 mm (0.25 in) diameter |
| Handle temperature range | 0 °C to 80 °C |
| Lead wires | 4-wire Super-Flex PVC, 22 AWG |
| Termination | Gold-plated spade lugs |
| Standard lengths | 12, 15 and 20 inches (304.8 mm, 381 mm, 508 mm) |
Source: Fluke 5626/5628 Platinum Resistance Thermometer User's Guide
All Fluke 5626/5628 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the difference between the Fluke 5626 and 5628 PRTs?
The two models use different standard resistance values: the 5626 is 100 Ω and the 5628 is 25.5 Ω.
From: Fluke 5626/5628 Platinum Resistance Thermometer User's Guide
What temperature range do the 5626 and 5628 cover?
They cover the International Temperature Scale of 1990 (ITS-90) range from −200 °C to 661 °C.
From: Fluke 5626/5628 Platinum Resistance Thermometer User's Guide
Can I use the 5626/5628 to measure the temperature of a live component?
No. The manual warns not to use the instrument to measure the temperature of any hazardous live component.
From: Fluke 5626/5628 Platinum Resistance Thermometer User Guide
