Start here

Fluke 9009 Industrial Dual-Block Calibrator User's Guide
The Fluke 9009 Industrial Dual-Block Calibrator User's Guide covers symbols, safety information, setup, power, temperature setting, display units, parts, and controls.
It also describes controller operation, serial communications, test probe calibration, calibration procedures, maintenance, and troubleshooting.
Fluke 9009 specifications
| Use environment | Indoor use only. |
|---|---|
| Installation category | OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1. |
| Well access temperature | Same as the actual temperature shown on the display. |
| Turn-off temperature limit | Do not turn off above 100°C. |
Source: Fluke 9009 Industrial Dual-Block Calibrator User's Guide
All Fluke 9009 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
Can the Fluke 9009 be used outdoors?
No. The guide states that the instrument is intended for indoor use only.
From: Fluke 9009 Industrial Dual-Block Calibrator User's Guide
When should the Fluke 9009 be energized after humid storage?
It should be energized for a 2-hour dry-out period before it can be assumed to meet IEC1010-1 safety requirements.
From: Fluke 9009 Industrial Dual-Block Calibrator User's Guide
What is the warranty period for the Fluke 9009?
The warranty period is one year and begins on the date of shipment.
From: Fluke 9009 Industrial Dual-Block Calibrator User's Guide
What is the maximum temperature at which the unit should not be turned off?
Do NOT turn off the unit at temperatures higher than 100°C.
From: Fluke 9009 Industrial Dual-Block Calibrator User's Guide