Key facts
- Brand
- Emerson
- Model
- 3144P
- Documents
- 2
- Languages
- EN · ZH
- Last added
- 2023-10-29
Start here

Emerson Rosemount 3144P Temperature Transmitter Quick Start Guide
This guide provides basic installation instructions for the Rosemount 3144P Temperature Transmitter. It covers system readiness, configuration verification, switch setup, mounting, wiring, and loop testing. Safety warnings for hazardous environments and compliance with explosion-proof standards are included. The guide references the full Reference Manual for detailed configuration and diagnostics. It also addresses HART revision compatibility and software requirements for Field Communicators.
Read onlineEmerson 3144P specifications
| Communication protocols | 4‑20 mA/HART and FOUNDATION Fieldbus (FF) |
|---|---|
| Power supply voltage | 12.0 V to 42.4 V DC |
| Input/Output isolation | 500 V peak, 707 V DC |
| Weight | 3.1 lb (1.4 kg) or 7.8 lb (3.5 kg) |
| Enclosure protection | IP66, IP68, NEMA 4X, CSA 4X |
| Operating temperature range | -40 °C to 85 °C (‑40 °F to 185 °F) |
Source: Emerson Rosemount 3144P Temperature Transmitter Manual
All Emerson 3144P documents by type
The shortest path to a working setup, usually only a few pages.
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the minimum power requirement for the Rosemount 3144P Transmitter?
Do not operate when power is below 12 Vdc at the transmitter terminal.
From: Emerson Rosemount 3144P Temperature Transmitter Quick Start Guide
What is the required loop resistance for communication with the transmitter?
The loop resistance must be between 250 and 1100 ohms.
From: Emerson Rosemount 3144P Temperature Transmitter Quick Start Guide
What communication protocols does the 3144P support?
It supports 4‑20 mA/HART and FOUNDATION Fieldbus (FF) communication.
From: Emerson Rosemount 3144P Temperature Transmitter Manual
What is the required power supply voltage for the 3144P?
The transmitter operates from a DC supply of 12.0 V to 42.4 V.
From: Emerson Rosemount 3144P Temperature Transmitter Manual