Thermo Scientific AquaClear low‑range turbidity analyzer product leaflet with specifications, features and ordering information.
The AquaClear low‑range turbidity analyzer from Thermo Scientific is designed for high‑precision, low‑cost turbidity monitoring in drinking water, membrane processes, bottled water and other water‑quality applications. Its compact 135 mL flow cell removes bubbles and provides stable readings with a long‑life white‑light source.
The leaflet includes technical parameters, optional accessories, ordering codes and typical industry applications such as municipal water, industrial pure water, and wastewater monitoring. This manual is written in Chinese.
| Measurement range | 0–100 NTU |
|---|---|
| Resolution | 0.001 NTU |
| Accuracy | ±2 % or ±0.015 NTU (whichever is larger) |
| Response time | 90 s to reach 90 % of final value |
| Analog output | 2 × 4–20 mA |
| Relay output | 2 × Form C, 5 A @250 VAC |
| Operating temperature (water) | -5 °C to 50 °C |
| Ambient temperature | -20 °C to 60 °C |
| Maximum flow rate | 500 mL/min |
| Minimum flow rate | 250 mL/min |
| Light source | White light (tungsten) |
| Sensor material | Quartz, anodized aluminum |
| Housing protection rating | IP66 (NEMA 4X) |
| Weight | 2.5 kg |
It measures turbidity from 0 to 100 NTU.
The instrument provides two analog output channels, each 4–20 mA.
The response time is 90 seconds to reach 90 % of the final value.
The housing is rated IP66 (NEMA 4X).