This manual provides basic operating instructions, parameter settings, safety notes, and maintenance for the Thermo Scientific Megafuge 16/40 and 16R/40R…
The Megafuge 16/40 (room‑temperature) and 16R/40R (refrigerated) are high‑capacity, multi‑function tabletop centrifuges designed for laboratory use. The device features an auto‑lock rotor system and a temperature‑controlled chamber.
The manual covers step‑by‑step operation, how to set acceleration, braking, speed, time, temperature, bucket selection, and program storage. It also includes safety precautions, cleaning and maintenance procedures, and a list of error codes with troubleshooting guidance. This manual is written in Chinese.
| Maximum speed | 15200 rpm |
|---|---|
| Minimum speed | 300 rpm |
| Temperature range | -10 °C to +40 °C (16R/40R only) |
After setting the desired parameters, press the START button; the machine will run until the time reaches zero and then stop automatically.
Press the TEMP key, then use the up/down arrows to adjust the temperature within the -10 °C to +40 °C range.
The maximum speed is 15200 rpm (equivalent to 25830 × g).
Rinse the rotor with distilled water and dry at room temperature; avoid soap, phosphoric acid, or bleach.