Rosemount 8700 Series electromagnetic flowmeter product catalog with selection, sizing, ordering information, specifications, certificates and dimension…
Rosemount 8700 Series product catalog (document 00813-0106-4727, Rev UB, May 2010) covers electromagnetic flowmeter selection, sizing, ordering information, specifications, product certificates, and dimension drawings. It presents transmitter models 8732E, 8712E, and 8712H together with sensor models 8705, 8707, 8711, and 8721.
The catalog lists a standard reference accuracy of 0.25% of reading and an optional high accuracy of 0.15% of reading. It describes transmitter output options including HART/analog, FOUNDATION Fieldbus, and Profibus PA, and gives sensor sizes and designs for flanged, wafer, large-signal, and sanitary applications. This manual is written in Chinese.
| Standard reference accuracy | 0.25% of reading |
|---|---|
| Optional high accuracy | 0.15% of reading |
| 8705 sensor size range | 1/2 in. (15 mm) to 36 in. (900 mm) |
| 8711 sensor size range | 0.15 in. (4 mm) to 8 in. (200 mm) |
| 8721 sensor size range | 1/2 in. (15 mm) to 4 in. (100 mm) |
| AC power supply (8732E/8712E) | 90 to 250 V AC, 50-60 Hz |
| DC power supply (8732E/8712E) | 12 to 42 V DC |
| Output options (8732E) | HART/analog, FOUNDATION Fieldbus, or Profibus PA |
The standard reference accuracy is 0.25% of reading; optional high accuracy is 0.15% of reading.
The 8705 sensor covers 1/2 in. (15 mm) through 36 in. (900 mm).
The AC power option is 90 to 250 V AC, 50-60 Hz.
It provides HART/analog, FOUNDATION Fieldbus, or Profibus PA output.