Rosemount 3051G pressure transmitter quick installation manual, mainly introduces the installation, jumper setting, wiring power supply, parameter setting and range adjustment of the transmitter
Rosemount 3051G pressure transmitter quick installation guide gives basic installation steps for the 3051G series. It covers mounting, jumper settings, wiring and power, parameter setup, range adjustment, and product approvals.
The guide also includes safety warnings for explosive and electrical hazards and states that full configuration, diagnostics, maintenance, troubleshooting, and explosion-proof installation details are in the Rosemount 3051 or 2088 product reference manuals. This manual is written in Chinese.
| Output signal | 4-20 mA |
|---|---|
| Power supply | 24 V DC |
| Power supply ripple | Less than 2% |
| Process connection | 1/2-14 NPT internal thread |
| Cable | 24 AWG or higher |
| Maximum cable length | 5,000 ft (1500 m) |
| Loop maximum resistance | 45.4 × (supply voltage - 10.5) ohms |
If the alarm or security jumpers are not installed, alarm defaults to high and write protection defaults to off.
Use shielded twisted pair.
Do not exceed 5,000 ft (1500 m).
Zero trim compensates for mounting position effects; use the HART communicator if the offset is no more than 3%, or the transmitter zero adjustment button if it exceeds 3%.