This manual describes the installation process for the Micro Motion 2200S Series transmitter
This is the Chinese-language installation manual (P/N MMI-20012962, Rev. BA, February 2009) for the Micro Motion 2200S transmitter. It describes the procedures needed to install the transmitter, including preparation, model code explanation, mounting options (integral or extended), and positioning of the transmitter and user interface module.
The manual also covers installation considerations such as power requirements and barrier requirements, wiring with and without a safety barrier or Micro Motion adapter barrier, and an appendix of dimensions and specifications. Further appendices describe the Micro Motion adapter barrier and the return policy. This manual is written in Chinese.
| Enclosure | NEMA 4X (IP67); cast aluminum with polyurethane paint or 316L stainless steel |
|---|---|
| Conduit connections | 1/2" NPT or M20 |
| Mounting | Integral or extended mounting; can be rotated on the mounting bracket in 45° increments for eight positions |
| Power supply | DC 17–36 VDC, loop resistance up to 600 Ω, maximum power 0.8 W |
| Terminal wire size | 26–14 AWG (0.14–2.5 mm²), solid or stranded |
| Channel A output | One passive 12–20 mA output; maximum load 600 Ω; external supply 17–36 VDC |
| Digital communication | HART Bell 202 at 1.2 and 2.2 kHz, 1200 baud, one stop bit, odd parity; requires 250–600 Ω resistance |
| Ambient temperature limits | Operating and storage: –40 to +140 °F (–40 to +60 °C) |
| Humidity limits | 5–95% relative humidity, non-condensing at 140 °F (60 °C) |
| Vibration limits | IEC 8.2.6, endurance sweep, 5–2000 Hz, 1.0 g, 50 sweep cycles |
H means the transmitter is supplied without a Micro Motion adapter barrier, and K means it is supplied with a Micro Motion adapter barrier.
It requires DC 17–36 VDC with loop resistance up to 600 Ω and a maximum power of 0.8 W.
Without an adapter barrier or third-party safety barrier the external device receives a 12–20 mA signal; with the Micro Motion adapter barrier installed it receives a 4–20 mA signal.
Operating and storage temperature range is –40 to +140 °F (–40 to +60 °C).