Installation Instructions for Micro Motion™ T Series Coriolis Flow and Density Sensor
This manual provides installation instructions for the Micro Motion T Series Coriolis flow and density sensor. It covers planning, including a checklist and best practices for mounting and temperature limits, as well as sensor and transmitter installation steps. The guide also details wiring options for 4-wire and 9-wire cables and grounding procedures. This manual is written in Japanese.
| Maximum cable length (Other MVD transmitters) | 18 m |
|---|---|
| Maximum power wire length (0.326 mm²) | 91 m |
| Maximum power wire length (0.518 mm²) | 152 m |
| Maximum power wire length (0.823 mm²) | 305 m |
| Maximum signal wire length | 305 m |
No straight pipe length is required upstream or downstream of the sensor.
Electronic components cannot be used if the ambient temperature is -40.0 °C or lower, or 60.0 °C or higher.