This manual is a guide for the use of the Model 5100 Analyzer
The AMETEK Model 5100 Analyzer uses tunable diode laser spectroscopy to measure gas concentrations in industrial processes. This manual provides comprehensive instructions for installation, start-up, operation via the user interface and web interface, and maintenance procedures.
It includes detailed safety notes covering electrical safety, grounding, sample gas hazards, warning labels, environmental compliance (WEEE), secondary process seal requirements, and electromagnetic compatibility guidelines. Specifications, menu maps, Modbus registers, and maintenance steps for the GRIN lens and mirror are also documented.
| Maximum secondary seal process pressure | 10 psig (170 kPa) |
|---|---|
| Maximum secondary seal process temperature | 140 °F (60 °C) |
| Hazardous location rating | Class I, Division 2, Groups A, B, C, D |
The maximum process gas pressure for the secondary seal is 10 psig (170 kPa).
The analyzer is suitable for use in Class I, Division 2, Groups A, B, C, D or non-hazardous locations only.
Instrument grounding is mandatory; performance specifications and safety protection are void if operated from an improperly grounded power source.