User manual for AMETEK's LG Series mechanical force gauges, covering operation, safety, peak hold, mounting, and specifications.
This manual provides instructions for operating and maintaining the AMETEK LG Series mechanical force gauges. It includes safety warnings, operational procedures for tension and compression testing, peak hold functionality, mounting guidelines, and specifications such as accuracy and overload capacity. The document also lists accessories and spare parts available for the gauge series.
| Accuracy | +0.5% of Full Scale |
|---|---|
| Overload Capacity | 110% of Rated Capacity |
| Maximum Overload | 150% of Rated Capacity |
| Tare Capacity | 10% of Rated Capacity |
| Dial Resolution | 0.5% of Full Scale |
| Deflection | 10mm (.394 inches) at Full Scale |
| Operating Temperature | +40°F to +110°F (+5°C to +45°C) |
| Instrument Weight | 1.5 lbs (0.67 kg) |
| Warranty | 2 Years |
Rotate the bezel on the dial indicator so that the indicator pointer lines up with the zero on the dial face.
The maximum overload capacity is 150% of the rated capacity.
No, the gauge is unidirectional and must be used with clockwise rotation only; counterclockwise rotation may cause damage.