This manual provides instructions for use of the Type T hydraulic deadweight testers. The manual covers product features, operational methods, and maintenance and care.
This manual provides step-by-step instructions for operating, calibrating, and maintaining AMETEK Type T hydraulic deadweight testers. It covers setup for single and dual column configurations, pressure generation, weight addition, calibration, and component maintenance. The device is designed for high-accuracy pressure testing, with options for different seal materials and test fluids. Special care is required for piston and cylinder assemblies due to their precision tolerances.
The manual includes troubleshooting for common issues like failure to pump or pressure loss, along with details on spare parts, fitting kits, and recertification procedures. It also references calibration data traceable to NIST and provides part numbers for included components and optional accessories.
| Accuracy | +/- 0.100% of output pressure |
|---|---|
| Pressure Range | 1 to 30 bar / 10 to 500 psi through 7 to 1000 bar / 100 to 15 000 psi |
| Test Fluids | Oil or water/alcohol mixture |
| Seal Packages | Buna-N, Viton®, and EPT |
Guaranteed accuracy is +/- 0.100% of output pressure, optionally 0.025% or 0.015%.
Models with "T" prefix use hard, non-magnetic die cast alloy; "TQ" models use stainless steel conforming to NIST Class Q.
Check for fluid contamination, foreign particles, or binding/freezing in the piston-cylinder annulus space.