AMETEK Type T Hydraulic Deadweight Tester Operation Manual

This manual provides instructions for use of the Type T hydraulic deadweight testers. The manual covers product features, operational methods, and maintenance and care.


AMETEK Type T Hydraulic Deadweight Tester Operation Manual - cover page
Brand
Ametek
Document type
User Manual
Language
English
Pages
43
File format
PDF
File size
2.1 MB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
3EBDE6B6065F0FEBD8AC05E22767DC68

About this manual

This manual provides step-by-step instructions for operating, calibrating, and maintaining AMETEK Type T hydraulic deadweight testers. It covers setup for single and dual column configurations, pressure generation, weight addition, calibration, and component maintenance. The device is designed for high-accuracy pressure testing, with options for different seal materials and test fluids. Special care is required for piston and cylinder assemblies due to their precision tolerances.

The manual includes troubleshooting for common issues like failure to pump or pressure loss, along with details on spare parts, fitting kits, and recertification procedures. It also references calibration data traceable to NIST and provides part numbers for included components and optional accessories.

What's inside

  • Introduction
  • Accuracy
  • Care and Handling
  • Features and Parts Lists
  • Setup
  • Operation
  • Maintenance and Replacement
  • Specifications
  • Support
  • Troubleshooting

Specifications

Accuracy+/- 0.100% of output pressure
Pressure Range1 to 30 bar / 10 to 500 psi through 7 to 1000 bar / 100 to 15 000 psi
Test FluidsOil or water/alcohol mixture
Seal PackagesBuna-N, Viton®, and EPT

Frequently asked questions

What is the accuracy of the Type T Hydraulic Deadweight Tester?

Guaranteed accuracy is +/- 0.100% of output pressure, optionally 0.025% or 0.015%.

What materials are the suspended weights made of?

Models with "T" prefix use hard, non-magnetic die cast alloy; "TQ" models use stainless steel conforming to NIST Class Q.

What should I do if the pump does not hold pressure?

Check for fluid contamination, foreign particles, or binding/freezing in the piston-cylinder annulus space.