Agilent 1290 Infinity Quaternary Pump Technical Overview

This Technical Overview demonstrates that the Agilent 1290 Infinity Quaternary Pump eliminates the need to premix the mobile phase, provides high-precision retention times and accurate quantification


Agilent 1290 Infinity Quaternary Pump Technical Overview - cover page
Brand
Agilent
Category
Other
Document type
Other
Language
English
Pages
4
File format
PDF
File size
804 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
22AA2937C0EB432F0B91FC6B1F4B2765

About this manual

This document presents a technical overview demonstrating that the Agilent 1290 Infinity Quaternary Pump eliminates the need to premix mobile phases. It compares the precision of retention times achieved with dynamically-mixed mobile phases against those obtained using premixed phases in ultrahigh performance liquid chromatography (UHPLC).

What's inside

  • Abstract
  • Introduction
  • Instrumentation and software
  • Chromatographic conditions
  • Results and discussion
  • Conclusions

Specifications

Pump ModuleAgilent 1290 Infinity Quaternary Pump (G4204A)
Autosampler ModuleAgilent 1290 Infinity Autosampler with Thermostat (G4226A, G1330B)
Column CompartmentAgilent 1290 Infinity Thermostatted Column Compartment (G1316C)
Detector ModuleAgilent 1260 Infinity Diode Array Detector (G4212A)
Flow rate1 mL/min
Column temperature30°C
Detection wavelength254/10 nm

Frequently asked questions

Does the Agilent 1290 Infinity Quaternary Pump require premixed mobile phases?

No, the pump eliminates the need to premix the mobile phase by providing highest precision using dynamically-mixed mobile phases.

How does dynamically-mixed mobile phase precision compare to premixed phases over several days?

Dynamically-mixed mobile phases showed better precision over several sequences and days compared to premixed phases prepared by the same or different users.

Which channels are used for dynamically-mixed mobile phases?

Channels A and B are used for dynamically-mixed mobile phases, while the C channel is deployed for premixed phases.