Agilent 6890/G2350A Stack Flare Gas Emissions Analysis Application Note

This document introduces a single-column technique for analyzing carbon, sulfur, and nitrogen compounds in a gaseous matrix. The analysis is performed with an Agilent 6890 gas chromatograph (GC) and an Agilent G2350A atomic emission detector (AED). An Agilent 30 m × 0.53 mm × 5 µm J&W DB-1 capillary column is used for separation.


Agilent 6890/G2350A Stack Flare Gas Emissions Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
105 KB
Published
03 July, 2012
Updated
30 September, 2023
Model
G2350A
MD5 checksum
4CC5449B9D5FB0953F510F0C34FA653B

About this manual

This document details an analytical method developed by Agilent Technologies to determine stack and flare gas emissions. The technique utilizes a single-column setup with an Agilent 6890 gas chromatograph coupled to a G2350A atomic emission detector (AED) to analyze carbon, sulfur, and nitrogen compounds in a gaseous matrix. The method addresses the need for timely and accurate analysis of pollutants like hydrocarbons, sulfur oxides, and nitrogen oxides from sources such as refineries and wastewater treatment plants.

What's inside

  • Introduction
  • Experimental
  • Carrier gas
  • Volatiles inlet
  • Table 1 GC Conditions
  • Results and Discussion
  • Conclusions
  • References

Specifications

Gas ChromatographAgilent 6890
DetectorAgilent G2350A AED
Column30 m x 0.53 mm x 5.0 um J&W DB-1
Oven Initial Temperature-46 °C
Oven Ramp Rate8 °C/min to 50 °C
Oven Final Temperature296 °C
Column Pressure3 psi

Frequently asked questions

What instruments are used for the stack gas analysis?

The analysis is performed with an Agilent 6890 gas chromatograph and an Agilent G2350A atomic emission detector.

What column is used for separation?

A 30 m x 0.53 mm x 5.0 um J&W DB-1 capillary column is used.

How many injections are required to quantify all components?

Three injections are required: one for hydrogen, one for carbon/sulfur/nitrogen, and one for oxygen-containing components.