This paper introduces the method of using UV-Vis spectroscopy to screen environmental samples.
Environmental regulations require routine analysis of complex field samples like soil and sludge, which often contain background contaminants that can interfere with analytical instruments. This document presents a study on using UV-Visible spectroscopy as a fast, non-destructive screening tool to identify samples likely to cause instrument contamination or low surrogate recovery during Gas Chromatography (GC) analysis.
The study utilized an Agilent 8453 system controlled by ChemStation software with automatic liquid sampling. Results demonstrated that the UV-Vis method could reliably detect harmful levels of hydrocarbons and predict surrogate recovery problems in samples containing PCBs, often identifying issues faster than dedicated GC screening methods.
| Analysis time per sample | 45 seconds |
|---|---|
| Sample volume range | 1 to 2 mL |
| Solvent flush volume | 4 to 6 mL |
| Sample recovery rate | 80 to 90 percent |
To identify environmental samples that are likely to cause problems during analysis, such as instrument contamination or low surrogate recovery.
Yes, the study found that UV-Vis successfully identified harmful levels of hydrocarbon even when gas chromatograph results showed none.
The study used an Agilent 8453 system controlled with Agilent's ChemStation software.