Agilent 1100 Series Improved 2D Nano LC/MS Application Note

This Application Note describes an improved method for protein identification with two-dimensional nano LC/MS, which utilizes the elution of digest peptides from a strong cation exchange column by an injected salt step gradient of increasing salt concentration with a subsequent reversed phase separation. However, in this approach the ion exchange chromatography does not produce its full performance. To improve the SCX chromatography performance a new method was developed, which works with a semi-continuous pumped salt gradient. This Application Note describes the improved method for two-dimensional nano LC/MS. To show the full performance of the method a complex tryptic digest of the yeast proteome was analyzed. The results obtained with the developed method are compared to the formerly used injection method as well as to an off-line 2D LC method.


Agilent 1100 Series Improved 2D Nano LC/MS Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
252 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
2603ABE8E46C77EB0CAD857741662EC8

About this manual

The document introduces an improved method for two-dimensional nano LC/MS that utilizes a semi-continuous pumped salt gradient to optimize strong cation exchange (SCX) chromatography performance. This approach addresses limitations of traditional injected salt plugs, such as poor peptide distribution and ionization suppression, by keeping the SCX column near equilibrium.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Experimental
  • Principle of two-dimensional nano LC-MS/MS
  • Control of pumps and valves
  • Chromatographic method

Frequently asked questions

How does the improved method differ from the traditional approach?

The improved method uses a semi-continuous pumped salt gradient instead of injected salt solution plugs to optimize SCX chromatography performance.

What is the purpose of this application note?

This document describes an improved method for two-dimensional nano LC/MS that works with a semi-continuous pumped salt gradient to improve SCX chromatography performance.

What sample was analyzed to demonstrate the method's performance?

A complex tryptic digest of the yeast proteome was analyzed to show the full performance of the developed method.

What is the starting concentration of the salt gradient for the first step?

The salt gradient begins at 0% to 2.5% of a 500 mM NaCl solution for the first step.