This document describes the method for using the Agilent 2100 bioanalyzer to perform flow cytometric analysis of a limited number of cells.
The document details how to perform flow cytometric analysis using a small number of cells with the Agilent 2100 bioanalyzer. It explains that while traditional methods require high cell volumes, this microfluidic system allows for analysis with as few as 2,500 cells per sample by utilizing on-chip staining protocols.
The document states that loading 2,500 cells onto the chip is sufficient for appropriate quantitation.
After vortexing, the chips must be stacked and incubated in the dark for 20 minutes at room temperature.
The recommended number of cells for this purpose is 20,000 cells per sample for most Agilent 2100 bioanalyzer cell applications.