Agilent ZORBAX Poroshell 300SB-C8 HPLC Columns Application Note: Use of Temperature to Increase Resolution in Ultrafast

This paper introduces a method for separating proteins using ZORBAX Poroshell 300SB columns under low pH conditions at high temperatures. The method uses a linear gradient elution method with a flow rate of 1.0 mL/min and a column temperature of 40 °C to 75 °C. The results show that the resolution of the peaks is significantly improved as the column temperature is increased.


Agilent ZORBAX Poroshell 300SB-C8 HPLC Columns Application Note: Use of Temperature to Increase Resolution in Ultrafast  - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
440 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
99318A5898F0A5C917ADDFBDDB64F70B

About this manual

This document describes an HPLC method that utilizes elevated temperatures to increase separation selectivity and resolution for complex peptide and protein mixtures. The study focuses on the stability of ZORBAX Poroshell 300SB columns at high temperatures, allowing them to be used effectively up to 90 °C.

What's inside

  • Introduction
  • Results
  • Highlights
  • Conclusions
  • Instrumentation

Specifications

Maximum Operating Temperature90 °C at low pH
Particle TypeSuperficially porous particles

Frequently asked questions

What is the maximum temperature for ZORBAX Poroshell 300SB columns at low pH?

The columns are stable and can be used at temperatures up to 90 °C.

How does temperature affect protein separation resolution?

Elevating the column temperature can effectively increase separation selectivity and resolution for ionizable compounds like proteins.

Download PDF (440 KB)