Agilent Technologies 6890/5973 Determination of Alkylphenols in Sewage Sludge Application Note

This document describes a routine method for the quantitative determination of 4-nonylphenol (NP), 4-nonylphenol monoethoxylate (NP1EO), and 4-nonylphenol diethoxylate (NP2EO) in sludge samples. A Soxtec® extraction procedure was used to enrich the target compounds from the solid matrix. Quantitative determinations were performed by high-speed gas chromatography/mass spectrometry using a short apolar fused silica column. Derivatization allows NP1EO and NP2EO to be analyzed by gas chromatography. The relative standard deviation was close to 5% for the analysis of 10 different sludges analyzed seven times each. Recoveries were determined for a sludge reference material and were higher than 90%. The experimental limits of quantification were 2, 5, and 5 µg/g of dry matter (µg/g DM), respectively, for NP, NP1EO, and NP2EO. Good agreement was observed between results obtained with this method and those obtained by our previous method, which used normal-phase liquid chromatography with an aminosilica column. This method was applied to the determination of alkylphenols and alkylphenol mono- and diethoxylates in sewage sludge by high-speed gas chromatography/mass spectrometry.


Agilent Technologies 6890/5973 Determination of Alkylphenols in Sewage Sludge Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
508 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
0EA94B560781C918C9473A82AEEB447A

About this manual

The document outlines a specific procedure for the quantitative determination of 4-nonylphenol (NP), 4-nonylphenol monoethoxylate (NP1EO), and 4-nonylphenol diethoxylate (NP2EO) in sludge samples. The method utilizes Soxtec extraction followed by derivatization with Sigma-Sil-A to enable analysis via gas chromatography/mass spectrometry.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Chemicals and Reagents
  • Sample Collection
  • Extraction Procedures
  • Instrumental Conditions
  • Tuning the Mass Selective Detector (MSD)
  • Method Validation

Specifications

GC/MS SystemAgilent 6890/5973 MSD
Column Type20-m x 0.10-mm id x 0.10 um RTX5 capillary column
Inlet Temperature300 °C
Source Temperature250 °C
Quadrupole Temperature150 °C
Transfer Line Temperature250 °C
Carrier Gas Flow0.3 mL/min

Frequently asked questions

What is the limit of quantification for NP, NP1EO, and NP2EO?

The experimental limits of quantification were 2, 5, and 5 µg/g of dry matter respectively.

What is the recovery rate for sludge reference material?

Recoveries determined for a sludge reference material were higher than 90%.

Which ions are monitored in SIM mode for NP1EO?

The monitoring ions chosen for NP1EO in SIM mode are 251, 265, and 336.