Agilent 4500 ICP-MS Semiquantitative Analysis of Glass Fragments Application Brief

This application brief describes a method for the semiquantitative analysis of glass fragments using laser ablation ICP-MS. The method can rapidly and accurately identify glass fragments, and requires no other chemical reagents.


Agilent 4500 ICP-MS Semiquantitative Analysis of Glass Fragments Application Brief - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
218 KB
Published
02 July, 2012
Updated
01 October, 2023
MD5 checksum
D0846C7F8E93E7D458AAFCA7249412EB

About this manual

The document details a method utilizing the Agilent 4500 ICP-MS and LSX-100 laser ablation system to characterize glass fragments. This approach provides rapid, semiquantitative elemental analysis that generates unique fingerprint patterns for sample identification with minimal sample preparation.

Key advantages of this technique include negligible sample consumption, which allows remaining samples for further testing, and the elimination of extensive digestion procedures required by traditional methods. The method relies on elemental ratios to create distinct fingerprints using specific elements such as aluminum, barium, calcium, iron, magnesium, manganese, strontium, titanium, and zirconium.

What's inside

  • Introduction
  • Experimental Methods
  • Instrumental Parameters
  • Sample Preparation
  • Elemental Ratios
  • Identification of Glass Samples
  • Reproducibility

Frequently asked questions

What elements were identified as providing a distinct elemental fingerprint for the glasses examined?

The nine elements identified are aluminum, barium, calcium, iron, magnesium, manganese, strontium, titanium and zirconium.

How much sample preparation is required for this analysis?

The whole sample preparation process was less than 3 minutes, involving washing with double distilled water, sonication for one minute, and drying with isopropyl alcohol.

Does this method require quantitative calibration of the ICP-MS instrument?

No, the use of elemental ratios eliminates the need for quantitative calibration of the ICP-MS instrument.