Agilent Capillary Electrophoresis System Anion Analysis Application Note

This document describes the method and results of analyzing anions in power plant waters using capillary electrophoresis. Anion analysis is performed using indirect UV detection, allowing for fast and reliable analysis of low ppb levels of small inorganic and organic anions.


Agilent Capillary Electrophoresis System Anion Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
80 KB
Published
02 July, 2012
Updated
29 September, 2023
MD5 checksum
825CE5C4FF6A66F41C9600C9A59A1314

About this manual

This document details a method for analyzing low levels of small inorganic and organic anions in power plant circuit waters using capillary electrophoresis (CE) with indirect UV detection. The technique is designed to monitor trace ionic contaminants to minimize operational costs and ensure safe plant operation.

What's inside

  • Abstract
  • Experimental and results
  • Chromatographic conditions
  • Precondition
  • Figure 1

Specifications

Detection limit1-5 ppb
Linearity range10-30 ppb
Injection voltage-5 kV
Capillary length58.5 cm
Effective length50 cm
Temperature30°C
Voltage30 kV
Detection wavelengths350/50 nm signal, 245/10 nm reference

Frequently asked questions

What is the detection limit for anion analysis?

The detection limit is 1-5 ppb.

What is the linear range of the assay?

The assay is linear over the range 10-30 ppb.

Which kit is used for the method?

The method uses the Agilent Inorganic Anion Solutions Kit (part number 5063-6511).