Agilent Monitoring of Electroless Plating Baths by Capillary Electrophoresis Application Note

The document is about the analysis method of electroless plating bath, using capillary electrophoresis technology, can detect metal ions and inorganic anions


Agilent Monitoring of Electroless Plating Baths by Capillary Electrophoresis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
108 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
C83568F4D34E2642FB11DF2AB869FF40

About this manual

This document details a method for monitoring electroless plating baths, specifically for non-metals like ceramics and plastics. It explains how capillary electrophoresis (CE) with indirect UV detection can effectively monitor metal cations, reducing agents, organic acids, and inorganic anions present in these bath solutions. The text highlights the advantages of CE over ion chromatography for analyzing complex bath compositions with minor sample preparation.

What's inside

  • Authors
  • Abstract
  • Experimental
  • Equipment
  • Results and discussion
  • Chromatographic conditions

Specifications

Detection limit1-2 ppm
Linearity range10-100 ppm
Repeatability (migration times)< 0.1 % RSD
Repeatability (peak area)< 4.5 % RSD
Capillary length80.5 cm total
Operating voltage-25 kV
Operating temperature20°C

Frequently asked questions

What kit is used for the analysis?

The analysis uses the Agilent Plating Bath Analysis Kit (Agilent part number 5064-8228).

How often should buffer vials be replaced?

Buffer vials should be replaced after 10 runs when using 2 mL-vials and after 5 runs when using 1 mL-vials.

What is the sample preparation method?

Sample preparation consisted simply of dilution with water.