Agilent 7820A GC Application Brief for Formic Acid Determination

Agilent 7820A GC System can be used to simply determine formic acid in acetic acid


Agilent 7820A GC Application Brief for Formic Acid Determination - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
207 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
27518B990BE5543402FC20CC06E023FE

About this manual

This document presents an analytical method for determining formic acid in acetic acid using the Agilent 7820A Gas Chromatograph. The method utilizes a micro thermal conductivity detector (pTCD) and an Agilent J&W DB-FFAP capillary column to achieve better separation and repeatability compared to previous packed column methods.

What's inside

  • Highlights
  • Experimental
  • Analytical conditions
  • Results
  • Conclusions

Specifications

ColumnAgilent J&W DB-FFAP, 30 m x 320 um, 0.25 um
Carrier gasHe, Constant flow: 1.5 mL/min
Oven temperature program80 °C (3 min) 8 °C/min 150 °C (5 min)
Detector temperaturepTCD: 200 °C
Detection limit (S/N=6.8)0.05 wt%
Method precision (RSD)1.75%
Linearity (r2)0.9917

Frequently asked questions

What column is used for the analysis?

An Agilent J&W DB-FFAP capillary column (30 m x 320 um, 0.25 um) is used.

What detector is used in this method?

The method uses a micro thermal conductivity detector (pTCD).

What is the detection limit for formic acid?

The method has a signal-to-noise of 6.8 for a formic acid concentration of 0.05 wt%.