Agilent 1290 Infinity LC System Application Note for Screening Impurities in Fine Chemicals

The document describes the method of analyzing impurities in octyl-dimethyl-p-aminobenzoic acid using the Agilent 1290 Infinity LC system. The system is capable of operating with long, high resolution columns under conditions of rapid analysis, using low viscosity acetonitrile and higher viscosity methanol as solvent conditions.


Agilent 1290 Infinity LC System Application Note for Screening Impurities in Fine Chemicals - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
718 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
3E9288371F98BE0838BB4469CBC34993

About this manual

The document describes an analysis of octyl-dimethyl-p-aminobenzoic acid (OD-PABA) for the presence of impurities, such as para-aminobenzoic acid (PABA). The study utilizes the Agilent 1290 Infinity LC system equipped with ultra violet/visible Diode Array detection to separate and detect these compounds.

What's inside

  • Introduction
  • Sample Preparation
  • LC Method Details
  • Results and Discussion
  • Conclusions

Frequently asked questions

What is the primary purpose of this application note?

The document demonstrates how to use the Agilent 1290 Infinity LC System with UV/VIS Diode Array detection to analyze octyl-dimethyl-p-aminobenzoic acid for impurities.

What are the specific dimensions of the columns tested in this application?

The system was used with 3.0 and 2.1 mm id C18 columns, including a 3 mm x 50 mm column and a 2.1 mm x 150 mm column.

What is the maximum operating pressure limit of the system described?

The Agilent 1290 Infinity LC system has a pressure limit of approximately 18,000 psi.