Agilent 1100 Series Polybutadiene GPC Analysis Application Note

This paper describes the method of analyzing polybutadiene using gel permeation chromatography


Agilent 1100 Series Polybutadiene GPC Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
75 KB
Published
02 July, 2012
Updated
01 October, 2023
MD5 checksum
A974BBA8F98B64D278D6C5CE73BD620B

About this manual

This document details the analysis of polybutadiene using Gel Permeation Chromatography (GPC) to evaluate molecular weight data for quality control in tyre production. It highlights the importance of stable temperature control and solvent degassing for accurate results.

What's inside

  • Abstract
  • Method Performance
  • Conditions
  • Sample Preparation
  • Equipment

Specifications

Mobile phaseTetrahydrofurane (THF)
Flow rate1 ml/min
Oven Temp20 °C
UV detectorDAD 254/100 nm, reference 360/100 nm
Refractive index detectorHP 1047A
SoftwareAgilent ChemStation + polymer labs GPC software

Frequently asked questions

What is the precision of the molecular weight data?

The precision of weight average molecular weight (Mw) is less than 1%, and the precision of number average molecular weight (Mn) is less than 2%.

What is the required precision for the pump's flow rate?

The pump should be able to pump the selected flow rate with a precision of typically less than 0.15 percent.

What is the recommended oven temperature for the analysis?

The oven temperature for the analysis is set to 20 degrees Celsius.

How much sample is required for the analysis?

The sample preparation requires 33 mg of sample dissolved in 1 ml of THF.