Agilent 6890 Series Trace Hydrocarbon Impurities in Ethylene Propylene Application Note

This application note describes a method for using an Agilent 6890 gas chromatograph to determine trace (low ppm) levels of hydrocarbon impurities in high-purity ethylene and propylene.


Agilent 6890 Series Trace Hydrocarbon Impurities in Ethylene Propylene Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
274 KB
Published
02 July, 2012
Updated
29 September, 2023
MD5 checksum
072ED42912E39BCCDD196AE3DA102CF4

About this manual

The document details a method to determine low ppm levels of hydrocarbon impurities in high-purity ethylene and propylene. It utilizes an Agilent 6890 Series gas chromatography system equipped with a heated gas sample valve, split/splitless inlet, and flame ionization detector (FID). The separation is achieved using an HP-PLOT Al2O3 column.

What's inside

  • Introduction
  • Experimental
  • Results and Discussion
  • Ethylene Analysis
  • Propylene Analysis
  • Summary

Frequently asked questions

What is the minimum concentration of hydrocarbon impurities that can be easily detected in ethylene and propylene?

Impurity levels below 10 ppm (mole) were easily detected in both ethylene and propylene.

What is the split ratio used for analyzing the propylene calibration standard shown in Figure 6?

The sample was analyzed at a split ratio of 20/1.