Agilent 1100 Series Analysis of Antioxidants and UV Stabilizers in Polymers

This document introduces the HPLC method of analyzing antioxidants and UV stabilizers.


Agilent 1100 Series Analysis of Antioxidants and UV Stabilizers in Polymers - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
177 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
4C139A2FD8B82DD540D83E7F05404576

About this manual

This document details a method for analyzing antioxidants and UV stabilizers in polymers using reversed-phase liquid chromatography with diode-array detection. The specific compounds analyzed include Uvinol 3000, Tinuvin P, Irganox 1098, Uvinol 3008, Lavinix BHT, Tinuvin 320, Irganox 1010, Irganox 1076, and Irgafos 168. The method utilizes a BDS column with a mobile phase of water containing tetrabutylammoniumhydrogensulfate and acetonitrile.

What's inside

  • Abstract
  • Conditions
  • Sample preparation
  • Method performance
  • Equipment
  • Formula of Antioxidants
  • Spectra of antioxidants

Specifications

Column125 x 3 mm BDS, 3 um
Mobile Phase AWater + 0.001 m Tetrabutylammoniumhydrogensulfate, pH = 3.0 with H,SO,
Mobile Phase BAcetonitrile
GradientStart with 30 % B, to 98 % Bin 10 min
Flow Rate0.5 ml/min
Injection Vol5
Oven Temp40 °C
UV-DetectorDAD, 280/20 nm
Reference900/50 nm
Limit of Detection<1ng
Precision of retention times0.2 %
Precision of areas< 3%

Frequently asked questions

What equipment is used for the analysis?

The analysis uses an Agilent 1100 Series system including a degasser, binary pump, autosampler, thermostatted column compartment, and diode array detector.

What software is used?

The system uses Agilent ChemStation software.

How are the samples prepared?

Polymer samples are dissolved in Tetrahydrofurane and filtered after extraction with an ultra-sonic bath for 30 min.