Agilent 6820 GC-NPD Analysis of Organophosphorus Pesticides Application Note

This document describes a method for studying fifteen of the most common and important organophosphorus pesticides (OPs) using the Agilent 6820 gas chromatograph (GC) equipped with the new nitrogen-phosphorus detector (NPD). The 6820 GC-NPD demonstrated good linearity for concentrations of pesticides range from 1 to 500 ng/mL (R2 >0.999) for most compounds. All OPs produced high signal-to-noise ratios for splitless injections at 10 ng/mL (ppb) concentrations with the NPD detector. Instrumental limits of detection for most OPs studied were at low or sub-ppb concentrations. This suggests the 6820 GC with an NPD is well suited for OP pesticide residue determinations in foods, water, or other samples.


Agilent 6820 GC-NPD Analysis of Organophosphorus Pesticides Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
473 KB
Published
12 June, 2012
Updated
29 September, 2023
Model
6820
MD5 checksum
6504C9D2DFE4C141389353A96AB88DD5

About this manual

The document describes the use of the Agilent 6820 gas chromatograph equipped with a new nitrogen-phosphorus detector (NPD) for studying fifteen common organophosphorus pesticides. The study demonstrates that the system provides good linearity and high signal-to-noise ratios, making it well-suited for determining pesticide residues in foods and water.

What's inside

  • Application
  • Introduction
  • Experimental
  • Instrumental Parameters
  • Results and Discussions
  • Conclusion

Specifications

Inlet temperature250°C
Injection modeSplitless
Carrier gasHe, head pressure: 12 psi
Detector temperature326 °C
Linearity range1 to 500 ng/mL

Frequently asked questions

What is the linearity range for the analyzed pesticides?

The system demonstrated good linearity for concentrations ranging from 1 to 500 ng/mL.

What solvent is recommended for the final injection solution?

It is best to use acetone or hexane, as solvents like methylene chloride can damage the detector.

Which software was used for instrument control and data processing?

Cerity Networked Data System (NDS) for Chemical QA/QC software was used.