The Agilent Technologies mass selective detector (MSD) coupled with deconvolution reporting software (DRS) provides additional powerful data processing capabilities to the MSD ChemStation software. Extensive data shown in this report supports the high confidence level that an analyst can have in results rapidly produced by the DRS.
The report details an application using the Agilent Technologies mass selective detector (MSD) coupled with deconvolution reporting software (DRS). The goal was to review full scan gas chromatography/mass spectrometry data for the confirmation of pesticide residues, a process that can be labor-intensive. The study compares the speed and accuracy of the automated DRS against manual analysis by experienced analysts.
The DRS is able to process a complex food extract total ion chromatogram in about 1 minute.
It can take an experienced analyst 15 to 30 minutes to review/confirm one data file.
Overall, 93% of the pesticides present in the Agilent RTL Pesticide database were detected and confirmed.