Identification of Unknown Pesticides in Food Using LC/MSD TOF and Ion Trap MS(n)

Identification of Unknown Pesticides in Food Using Both LC/MSD TOF and Ion Trap MSn


Identification of Unknown Pesticides in Food Using LC/MSD TOF and Ion Trap MS(n) - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
321 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
D0E18AA91A03A111453F3D3BB2DBB908

About this manual

The document outlines an identification scheme for unknown pesticides that combines liquid chromatography/mass selective detection (LC/MSD) with time-of-flight (TOF) and ion trap options. The approach involves accurate mass determination using LC/MSD TOF followed by MS/MS confirmation using the ion trap, and database analysis against Merck Index or ChemINDEX.

The procedure details a six-step process including initial detection of unknowns in vegetable extracts, checking for isotope patterns (A+2), searching databases for empirical formulas, performing ion trap MS/MS fragmentation analysis, combining data for tentative identification, and final confirmation with authentic standards. Two examples are provided where malathion and buprofezin were identified in cucumber and tomato extracts.

What's inside

  • Abstract
  • Introduction
  • Experimental Methods
  • Standard Vegetable Extraction
  • Rapid Vegetable Extraction
  • LC/MSD TOF Methods
  • LC/MSD Ion Trap Methods
  • Results and Discussion

Specifications

ColumnZORBAX Eclipse XDB-C8, 4.6 mm x 150 mm x 5 um
Flow rate0.6 mL/min
Injection volume50 uL
Mobile phase AACN

Frequently asked questions

How are sulfur-containing pesticides identified?

They are identified by checking for an A+2 ion pattern (approximately 8% intensity) and confirming with MS/MS fragmentation.

What is the maximum residue limit for banned compounds in vegetables?

The maximum residue limit for banned compounds is set at 0.01 mg/kg.

What are the dimensions of the ZORBAX Eclipse XDB-C8 column used?

The column dimensions are 4.6 mm x 150 mm with a particle size of 5 um.

What is the flow rate for the LC/MSD TOF and Ion Trap methods?

The flow rate used in both methods is 0.6 mL/min.