Screening for EU Banned Dyes in Textiles Using Agilent 1120/6140 Application Note

This article describes a method for screening for EU banned dyes in textiles using the Agilent 1120 Compact LC with an Agilent 6140 Single Quadrupole LC/MS system and the Analytical Studio Browser software. The method allows for quick identification of samples that fail regulatory requirements, increasing productivity.


Screening for EU Banned Dyes in Textiles Using Agilent 1120/6140 Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
1.4 MB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
78FE2C8CA7D65F29A14C7DA37A738398

About this manual

The document details a modified literature method for screening EU-banned azo dyes in textiles. The process involves reducing the dyes to aromatic amines, which are then extracted and analyzed using an Agilent 1120 Compact LC coupled to an Agilent 6140 Single Quadrupole LC/MS system.

The analysis utilizes Analytical Studio Browser (ASB) software to rapidly identify samples that exceed regulatory limits for carcinogenic aromatic amines. The method employs a specific gradient run on an Agilent ZORBAX Eclipse Plus C18 column and operates the mass spectrometer in time-programmed SIM mode.

What's inside

  • Introduction
  • Experimental Parameters
  • Positive Control
  • Results and Discussion
  • Analytical Studio Browser

Specifications

ColumnAgilent ZORBAX Eclipse Plus C18, 150 mm x 3.0 mm, 3.5 um
Flow rate0.7 mL/min
Injection volume5 uL
Column temperature30 C
Detection wavelength (VWD)254 nm

Frequently asked questions

What is the regulatory limit for aromatic amines in textiles?

The concentration should not exceed 30 ppm for each amine.

How many restricted aromatic amines are listed by the EU?

The EU has listed 22 aromatic amines as potential carcinogens.

What software is used for rapid identification of samples?

Analytical Studio Browser (ASB) allows quick visual identification of samples that exceed regulatory requirements.