Agilent Poroshell 120 Columns Application Note

Agilent Poroshell 120 columns offer the highest efficiency, while significantly reducing back pressure


Agilent Poroshell 120 Columns Application Note - cover page
Brand
Agilent
Category
Other products
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
574 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
7400117365F5212F0D5AA0B7536782FA

About this manual

This document is an application note by Agilent Technologies demonstrating the efficiency of Poroshell 120 superficially porous columns. It shows that coupling three 150 mm columns can achieve over 100,000 plates in less than 5 minutes while keeping back pressure manageable. The note compares these columns to sub-2-µm particle columns, highlighting lower back pressure and higher peak capacity.

What's inside

  • Abstract
  • Introduction
  • Experimental Equipment
  • Potential benefits of superficially porous columns
  • Configuring the system
  • Column efficiency evaluation
  • Precision of retention times
  • Comparison of peak capacity

Specifications

Particle size2.7 µm
Core size1.7 µm
Shell size0.5 µm
Back pressure reduction40-50% lower
Efficiency comparison80-90% of sub-2-µm

Frequently asked questions

What is the maximum back pressure for Agilent Poroshell 120 columns?

These columns can be used up to 600 bar.

How many plates can be obtained with three coupled columns?

Approximately 103,000 plates were obtained at 1.5 mL/min flow rate.

What is the peak capacity of the Poroshell 120 column compared to a sub-2-µm column?

The Poroshell 120 column showed 133 peaks, which is 30% higher than the sub-2-µm column's 101 peaks.